Example: Simple OSA Measurement
For this example on how to get started, we'll be looking at a simple measurement using an Optical Spectrum Analyzer (OSA). We will need to do two things:
- Implement the measurement in a new measurement class which we will call
ReadOSA
- and interface the OSA via the instrument driver class called
OpticalSpectrumAnalyzerAQ6370C
This how-to describes the coding of the measurement class. You can find the final file at LabExT/Measurements/ReadOSA.py. We encourage you to read this code file in parallel while we explain the various parts here.
Measurement File
The example measurement will have one purpose: To capture the optical spectrum as recorded at the OSA and return the data collected.
Lets start by creating a new file - ReadOSA.py
- in LabExT/LabExT/Measurements
. The file should include
the measurement class with the same name ReadOSA
. A raw skeleton is of the form
#!/usr/bin/env python3
# -*- coding: utf-8 -*-
"""
LabExT Copyright (C) 2021 ETH Zurich and Polariton Technologies AG
This program is free software and comes with ABSOLUTELY NO WARRANTY; for details see LICENSE file.
"""
from LabExT.Measurements.MeasAPI import *
class ReadOSA(Measurement):
def __init__(self, *args, **kwargs):
super().__init__(*args, **kwargs) # calling parent constructor
self.name = 'ReadOSA'
self.settings_path = 'ReadOSA_settings.json'
self.instr_osa = None
A lot of the needed functionality is already implemented in the Measurement
class, located at
LabExt/LabExT/Measurements/MeasAPI/Measurement.py
,
which the new measurement must subclass. A measurement must include the following methods:
- constructor (
__init__
): must call the parent constructor, define the measurement name (self.name
), settings path (self.settings_path
) and the instruments instances used in the measurement (in this case onlyself.instr_osa
). The instrument will then dynamically be set to an instance of the instrument class which we will implement later. - method
get_default_parameter
: returns the wanted parameters. - method
get_wanted_instrument
: returns the wanted instrument types. - method
algorithm
: conducts the actual measurement.
get_default_parameter
Takes no arguments and returns the default parameters. This method sets the types, units and default values of the parameters used by the measurement. An example is:
'OSA center wavelength': MeasParamFloat(value=1550.0, unit='nm')
LabExT/LabExT/Measuremens/MeasAPI/MeasAPI.py
, please see its documentation for an introduction.
Set the parameter as seen above: The default wavelength that the spectrum recorded at the OSA should be centered around
is (in our case) 1550 nm, thus we create a MeasParamFloat
instance with value of 1550.0 and unit of 'nm'.
For our case, we also want to indicate other parameters, such as span, the resolution bandwidth and the number of steps
on the wavelength axis. So, the final parameter section looks like:
@staticmethod
def get_default_parameter():
return {
# osa settings
'OSA center wavelength': MeasParamFloat(value=1550.0, unit='nm'),
'OSA span': MeasParamFloat(value=2.0, unit='nm'),
'no of points': MeasParamInt(value=2000),
'sweep resolution': MeasParamFloat(value=0.08, unit='nm')
}
This method does not require a class instance, so we make it a staticmethod.
Hint
Depending on the MeasParam
kind you choose, LabExT will offer the user different GUI elements for entering the
parameters: Float
, Int
and String
get a text box, Bool
gets a checkbox, List
gets a dropdown menu.
get_wanted_instrument
Takes no arguments and returns a list of the wanted instrument types. The instrument types are set in
~/.labext/instruments.config
,
which consists of a dictionary with key-value-pairs where the instrument types correspond to the keys and a list of the
available instruments to the value. Please see the configuration page for an introduction.
As this measurement requires only one instrument, we return the one-element list with just requiring the OSA
type.
@staticmethod
def get_wanted_instrument():
return ['OSA']
algorithm
Finally, we implement the measurement algorithm in the algorithm
method.
It takes the four arguments detailed below. This function will be called by the StandardExperiment
located in LabExT/LabExT/Experiments/StandardExperiment.py
when this measurement is run via the GUI.
- device: name of the device set in the GUI. Unused here.
- data: dictionary pre-filled with information set in the GUI. In the dict stored at
data['values']
we should add lists of our measured data vectors. In the dict atdata['measurement settings']
we should store the final used measurement parameters. See the data dict documentation for further details. - instruments: list of instruments, in this case a one-element dict consisting of an instance of the OSA class.
- parameters: dictionary of parameters, similar to the dict returned by
get_default_parameter
but filled with the values set via the GUI.
As a rule of thumb, the algorithm method implementation should follow these points:
- read necessary parameters and save them to local variables. Note that to get the parameter value you have to access
the
.value
member of the MeasParam class instance (MeasParamFloat
,MeasParamInt
, etc.). - write the measurement parameters into the measurement settings (will be saved into the result JSON file at the end).
- get instrument pointers and open instrument connections.
- set instrument parameters.
- run your measurement routine and receive data.
- save data.
- close instrument connections
- return data.
1. read parameters
For easy of use, we read out the parameters from the parameters
argument to local variables.
def algorithm(self, device, data, instruments, parameters):
# get osa parameters
osa_center_wl_nm = float(parameters.get('OSA center wavelength').value)
osa_span_nm = float(parameters.get('OSA span').value)
sweep_resolution_nm = float(parameters.get('sweep resolution').value)
no_points = int(parameters.get('no of points').value)
2. check and save parameters
Some measurements require some checks on the parameters, e.g. if a desired laser wavelength is outside of the achievable range of your actual laser instrument.
Here we don't need to do checks and can directly save the used parameters into the output dictionary data
. See
the data dict documentation for further details.
# write the measurement parameters into the measurement settings
for pname, pparam in parameters.items():
data['measurement settings'][pname] = pparam.as_dict()
3. open instrument connection
We then retrieve the instrument driver instance and open the network connection to it:
self.instr_osa = instruments['OSA']
self.instr_osa.open()
4. set instrument parameters
We can now communicate with the instrument, lets set its properties. We want to set the span, the center wavelength, the resolution bandwidth, and the number of points. As we use property-setters in the instrument driver, these lines actually will cause SCPI messages to be sent to the instrument.
# set instrument parameters
self.instr_osa.span = osa_span_nm
self.instr_osa.centerwavelength = osa_center_wl_nm
self.instr_osa.sweepresolution = sweep_resolution_nm
self.instr_osa.n_points = no_points
5. run measurement and read data
Finally we advice the instrument to trigger a new sweep and to read the data into the variables
x_data_nm
and y_data_dbm
:
# everything is set up, run the sweep
self.logger.info('OSA running sweep')
self.instr_osa.run()
sleep(0.5)
# pull data from OSA
x_data_nm, y_data_dbm = self.instr_osa.get_data()
self.logger.info('OSA data received')
The run()
method of the instrument driver triggers a new measurement on the OSA and then waits until
the wavelength sweep is completed. After a small wait, the get_data()
then retrieves the measured data from the instrument.
6. save the data
We must save the data in the data['values']
dictionary under appropriate names. See
the data dict documentation for further details.
# copy the read data over to the json
data['values']['wavelength [nm]'] = x_data_nm
data['values']['transmission [dBm]'] = y_data_dbm
7. close instrument connections
We tell the VISA library to close the connection to the instrument:
# close instrument connection
self.instr_osa.close()
self.logger.info('closed connection to OSA')
8. check and return data
To make sure that we have filled all required keys of the data
dictionary, we call a check and finally
return the data dict.
# check data for sanity
self._check_data(data)
return data
Docstring
To give the user of this measurement an idea on how to use this class, we also write the docstring for this measurement class. The docstring will automatically be parsed and shown in the help when you press F1 inside LabExT. You should include a basic description, an example laboratory setup for when this measurement could be helpful and a description of all parameters.
The docstring must be formatted in markdown. Ours looks like this:
## ReadOSA
This class performs a measurement that sets up and runs a single sweep at an OSA and returns the data collected.
Data returned corresponds to a 'snapshot' of the OSAs screen after a sweep.
### Example Setup
```
DUT -> OSA
```
The only instrument required for this measurement is an OSA, either a APEX, Yokagawa or HP model. This measurement
can for example be used in longterm measurements where a signal generator and laser are operated manually.
### OSA Parameters
- **OSA center wavelength**: Center wavelength of the OSA in [nm]. Stays fixed throughout the measurement if
autocenter is disabled.
- **auto center**: If selected, the OSA runs a sweep and determines the center wavelength by itself. If disabled,
then the wavelength set in 'OSA center frequency' is used.
- **OSA span**: Span of the OSA in [nm]. Determines in which region the spectrum is recorded. The optical power at
all wavelengths within [$f_{center} - \frac{span}{2}$, $f_{center} + \frac{span}{2}$] are recorded.
- **number of points**: Number of points the OSA records. Should be bigger than span / sweep resolution
- **sweep resolution**: Resolution of the OSA sweep in [nm]. Allowed values are dependent on the OSA model.
Yokagawa AQ6370C: 0.02 nm and 2 nm, HP70951A: > 0.08 nm, APEX: > 4e-5 nm.
This concludes the measurement class implementation for the ReadOSA
functionality.